Interpretation Answer
In this test we will be using the FEV6 as a surrogate for the FVC. The FEV6 is obtained when the maneuver is shortened by terminating the maneuver after only 6-seconds of exhalation. The FEV6 is not recommended for use in occupational testing.
The first step in interpreting results is to determine if you have a valid test. In this test, we do have three acceptable maneuvers, but we do not have a repeatable test. However, the lack of a repeatable test is not a sufficient condition to warrant failing to interpret the results -- particularly in patients with airways obstruction who have a greater difficulty in producing a repeatable test. In this case, the FEV1/FEV6% is well below the LLN which would suggest an obstructive pattern. However in these results, both the FEV6 and FEV1 are greater than 120% or the predicted values. Since 100% is the average percent predicted in a normal population, these results are not consistent with an obstructive lung disease pattern.

The explanation for these results is a flow zero
error for the following reasons:
1. The test in not repeatable, particularly FEV6
(FVC).
2. The percent predicted values (FEV1 and particularly FEV6
or FVC) are above 120% with
a low FEV1/FEV6
(FEV1/FVC%).
3. The volume is constantly increasing or there is a constant flow (see lines on
graphs). Even with airways obstruction, the volume-time curve should
eventually plateau. However, it is difficult to determine if the curve
would not have eventually exhibited a plateau when only a 6-second maneuver is
performed. Some individuals, particularly those with airways obstruction,
may require longer than 6-seconds to reach a plateau. This is one of
several reasons why the shortened maneuver of only 6-seconds
(FEV6)
is not recommended for use in occupational testing as a surrogate for
FVC.
Although the above criteria may occur without a flow-zero error, to see all three conditions in the same test is highly suggestive of a flow zero error. Zero Errors can occur due to faulty electronic zeroing of the pressure transducer, resulting in either a false steady increase in flow or a false steady decrease in flow much like a leak. Although it is dependent on the type of flow-spirometer, usually the faulty zeroing occurs when the sensor is moved or air is allowed to flow through the sensor while a zero is being determined.
The test should be repeated, taking greater care to keep the flow sensor still, without any air flowing through the sensor, while the instrument is performing a flow zero. Regardless of whether there is a flow-zero error, the lack of a repeatable test would suggest some caution in interpreting the results as a mild obstructive lung disease pattern.